CTS T-40/50 — Environmental Chamber

Overview

The CTS T-40/50 is an environmental test chamber for temperature and humidity testing. It uses a proprietary binary ASCII protocol over RS-232 — not VISA/SCPI.

Specification Value
Type Environmental Test Chamber
Temperature range −75 … +185 °C
Interface RS-232 (COM port), 19200 baud, 8N1 with odd parity
Protocol Custom binary ASCII framing (STX / ADR / DATA / CHK / ETX)
Driver name CTSChamber
Resource patterns COM?*

Serial Connection Settings

The CTS chamber does not use VISA. The driver uses a custom IConnectionProvider that opens a COM port directly:

Parameter Value
Baud rate 19200
Data bits 8
Parity Odd
Stop bits 1
Flow control None

Connecting

  1. Connect the CTS chamber's RS-232 port to the PC (use a USB-to-RS232 adapter if needed)
  2. Note the COM port number from Device Manager
  3. Click Add Instrument and select the COM port resource (e.g. COM3)
  4. The Connection Manager lists COM ports automatically
  5. Select the CTSChamber driver from the drop-down (auto-detection is not possible because the chamber does not respond to *IDN?)
  6. Click Connect

USB-to-RS232 adapters

Use a high-quality adapter with a genuine FTDI or Prolific chip. Cheap adapters often lose data at 19200 baud with odd parity.


Protocol Details

The CTS chamber uses a proprietary binary ASCII framing scheme. Developers who need to extend the driver should read the CTSSerialConnectionProvider documentation.

Frame structure:

STX  |  ADR  |  DATA (each byte OR'd with 0x80)  |  CHK  |  ETX
0x02    0x30    0xC2 0xD4 ...                        0xHH   0x03

The checksum (CHK) is the XOR of all bytes from ADR to the last DATA byte.


Front Panel

Control Description
Set Temperature (°C) Target temperature
Set Ramp (°C/min) Temperature ramp rate
Start Start the chamber program
Stop Stop the chamber
Pause Pause the current program
Temperature readback Current chamber temperature in °C
Status Chamber operating mode (Running, Paused, Idle, Error)

Sequence Blocks

CTSChamber_SetTemperature

Sets the target temperature set-point.

Property Type Description
Instrument InstrumentSelector Connected CTS chamber instance
Temperature (°C) Number Target temperature

CTSChamber_SetRamp

Sets the ramp rate between temperature set-points.

Property Type Description
Instrument InstrumentSelector Connected CTS chamber instance
Ramp Rate (°C/min) Number Rate of temperature change

CTSChamber_Start

Starts the chamber conditioning program.

Property Type Description
Instrument InstrumentSelector Connected CTS chamber instance

CTSChamber_Stop

Stops the chamber and returns to standby.

Property Type Description
Instrument InstrumentSelector Connected CTS chamber instance

CTSChamber_Pause

Pauses the current program without losing the set-point.

Property Type Description
Instrument InstrumentSelector Connected CTS chamber instance

CTSChamber_ReadTemperature

Reads the current chamber temperature.

Property Type Description
Instrument InstrumentSelector Connected CTS chamber instance
Variable Variable Receives temperature in °C

CTSChamber_WaitForTemperature

Blocks execution until the chamber reaches a temperature within a tolerance band, or a timeout expires.

Property Type Description
Instrument InstrumentSelector Connected CTS chamber instance
Target (°C) Number Target temperature
Tolerance (°C) Number Acceptable deviation band (e.g. 0.5 means ±0.5 °C)
Timeout (s) Number Maximum wait time in seconds. 0 = no timeout.

Example Sequence — Temperature Profile Test

[Start]
→ [CTSChamber_SetRamp, Rate=2.0]
→ [CTSChamber_Start]
→ [Loop temperatures in -40, 25, 85°C: use SetVariable]
    ↓
    [SetVariable, Name=targetTemp, Value=-40]
    → [CTSChamber_SetTemperature, Temp={targetTemp}]
    → [CTSChamber_WaitForTemperature, Target={targetTemp}, Tolerance=0.5, Timeout=3600]
    → [Wait 300000ms]   # soak 5 minutes
    → [HP34401A_MeasureDCV, Variable=voltage]
    → [SaveCsv, Path="C:\Data\temp_profile.csv", Append=true]
→ [CTSChamber_Stop]
→ [End]