Skip to content

Rigol DS1000Z — 4-Channel Oscilloscope

Overview

The Rigol DS1054Z and DS1104Z are 4-channel digital oscilloscopes with 50 MHz and 100 MHz bandwidth respectively. They connect via LAN (TCPIP), USB, or GPIB and communicate over standard VISA.

Specification Value
Type Digital Oscilloscope
Models DS1054Z (50 MHz), DS1104Z (100 MHz)
Channels 4 analog
Interface LAN (TCPIP), USB-TMC, GPIB
Driver name RigolDS1000Z
Resource patterns TCPIP?*::?*::INSTR, TCPIP?*::hislip?*::INSTR, USB?*::0x1AB1::?*::INSTR, GPIB?*::?*::INSTR

Connecting

  1. Connect the DS1000Z via LAN or USB
  2. Click Add Instrument and select the resource
  3. Auto-detection parses *IDN? (RIGOL TECHNOLOGIES,DS1054Z or DS1104Z)
  4. Click Connect

For LAN: find the instrument IP address in its Utility → IO Setting → LAN menu, then enter the resource manually in the Connection Manager: TCPIP0::192.168.x.x::inst0::INSTR


Front Panel

Control Description
CH1–CH4 enable Toggle to show/hide each channel
V/div per channel Vertical scale selector
Coupling DC / AC / GND per channel
Probe ratio ×0.1 … ×1000 per channel
BW limit OFF or 20 MHz bandwidth limit
Invert Invert signal polarity
Timebase Horizontal time/div selector
Timebase mode MAIN / XY / ROLL
Trigger mode EDGE / PULSe / SLOPe / VIDeo / PATTern / RS232 / I2C / SPI
Trigger source CH1–CH4 or AC
Trigger level Threshold in V
Trigger slope Positive / Negative / Either
Trigger sweep AUTO / NORMal / SINGle
Acquire type Normal / Averages / Peak Detect / Hi-Res
Run button Start continuous acquisition
Stop button Freeze the display
Single button One-shot acquisition
AutoScale button Auto-configure vertical and horizontal scales
Math / FFT Configure MATH channel and FFT analysis

Sequence Blocks

RigolDS1000Z_SetChannel

Configures a single analog channel.

Property Type Description
Instrument InstrumentSelector Connected DS1000Z instance
Channel Combo 1, 2, 3, 4
Display CheckBox Enable or disable the channel
Coupling Combo DC, AC, GND
V/div Combo Vertical scale (0.001 V … 100 V/div)
Offset [V] Text Vertical offset
Probe × Combo Probe attenuation ratio (0.1 × … 1000 ×)
BW limit Combo OFF or 20M (20 MHz limit)
Invert CheckBox Invert channel polarity

RigolDS1000Z_SetTimebase

Sets the horizontal timebase.

Property Type Description
Instrument InstrumentSelector Connected DS1000Z instance
s/div Combo Time per division (5 ns … 50 s)
Time offset [s] Text Horizontal offset from trigger point
Timebase mode Combo MAIN, XY, ROLL

RigolDS1000Z_SetTrigger

Configures the trigger system. Edge trigger parameters are applied when Trigger mode is EDGE.

Property Type Description
Instrument InstrumentSelector Connected DS1000Z instance
Trigger mode Combo EDGE, PULSe, SLOPe, VIDeo, PATTern, RS232, I2C, SPI
Sweep Combo AUTO, NORMal, SINGle
Source Combo CH1, CH2, CH3, CH4, AC
Slope Combo POSitive (rising), NEGative (falling), RFALl (either)
Level [V] Text Trigger threshold voltage

RigolDS1000Z_SetAcquire

Configures the acquisition mode and memory depth.

Property Type Description
Instrument InstrumentSelector Connected DS1000Z instance
Acquire type Combo NORMal, AVERages, PEAKdetect, HRESolution
Averages Combo 2, 4, 8 … 1024 (active when type = AVERages)
Memory depth Combo AUTO, 12 k, 120 k, 1.2 M, 12 M, 24 M points

RigolDS1000Z_Run

Starts continuous acquisition (RUN).

Property Type Description
Instrument InstrumentSelector Connected DS1000Z instance

RigolDS1000Z_Stop

Stops acquisition and freezes the display (STOP).

Property Type Description
Instrument InstrumentSelector Connected DS1000Z instance

RigolDS1000Z_Single

Arms the oscilloscope for a single-shot acquisition (SINGle), then waits.

Property Type Description
Instrument InstrumentSelector Connected DS1000Z instance
Wait after [ms] Number Delay after sending the command (default 500 ms)

RigolDS1000Z_AutoScale

Sends :AUToset to automatically configure vertical scale, timebase, and trigger.

Property Type Description
Instrument InstrumentSelector Connected DS1000Z instance
Wait after [ms] Number Delay to let the scope settle (default 2000 ms)

RigolDS1000Z_MeasureVoltage

Reads an automated voltage-domain measurement from one channel.

Property Type Description
Instrument InstrumentSelector Connected DS1000Z instance
Channel Combo 1, 2, 3, 4
Parameter Combo VMAX, VMIN, VPP, VTOP, VBASe, VAMP, VAVG, VRMS, OVERshoot, PREShoot
Output variable Variable Receives the measurement value in volts

RigolDS1000Z_MeasureTime

Reads a time-domain or frequency measurement from one channel.

Property Type Description
Instrument InstrumentSelector Connected DS1000Z instance
Channel Combo 1, 2, 3, 4
Parameter Combo FREQuency (Hz), PERiod (s), RISetime (s), FALLtime (s), PWIDth (s), NWIDth (s)
Output variable Variable Receives the measurement value

RigolDS1000Z_MeasureDuty

Reads the duty cycle of a periodic signal.

Property Type Description
Instrument InstrumentSelector Connected DS1000Z instance
Channel Combo 1, 2, 3, 4
Parameter Combo PDUTycycle (positive), NDUTycycle (negative)
Output variable Variable Receives the duty cycle in %

RigolDS1000Z_ReadWaveform

Downloads the raw waveform data from one channel and saves it as a CSV file with columns Czas[s] and Napięcie[V].

Property Type Description
Instrument InstrumentSelector Connected DS1000Z instance
Channel Combo 1, 2, 3, 4
Waveform mode Combo NORMal (screen points), MAXimum, RAW (full memory)
CSV file path FilePath Destination file for the waveform data

RigolDS1000Z_MathSetup

Configures the MATH channel for arithmetic operations or FFT analysis.

Property Type Description
Instrument InstrumentSelector Connected DS1000Z instance
Operation Combo ADD, SUBTract, MULTiply, DIVision, FFT, ANDer, ORer, XORer, NOTer, INTGral, DIFF, SQRT, LOG, LN, EXP, ABS
Source 1 Combo CH1, CH2, CH3, CH4
Source 2 Combo CH1, CH2, CH3, CH4 (ignored for unary operations)
Enable Math CheckBox Show or hide the MATH channel on screen
FFT: window Combo RECTangle, BLACkman, HANNing, HAMMing, FLATtop, TRIangle
FFT: unit Combo VRMS, DB

Example Sequence — Single-Shot Capture with Measurement

[Start]
→ [RigolDS1000Z_SetChannel, Channel=1, Coupling=DC, V/div=1, Probe×=10]
→ [RigolDS1000Z_SetTimebase, s/div=0.001, Mode=MAIN]
→ [RigolDS1000Z_SetTrigger, Mode=EDGE, Sweep=SINGle, Source=CH1, Slope=POSitive, Level=0.5]
→ [RigolDS1000Z_Single, WaitMs=2000]
→ [RigolDS1000Z_MeasureVoltage, Channel=1, Parameter=VPP, OutputVariable=vpp]
→ [RigolDS1000Z_MeasureTime, Channel=1, Parameter=FREQuency, OutputVariable=freq]
→ [LogMessage "VPP={vpp} V  Freq={freq} Hz"]
→ [RigolDS1000Z_ReadWaveform, Channel=1, WaveMode=NORMal, FilePath=C:\Data\waveform.csv]
→ [SaveCsv, Path=C:\Data\results.csv]
→ [End]

Example Sequence — FFT Analysis

[Start]
→ [RigolDS1000Z_SetChannel, Channel=1, Coupling=DC, V/div=0.5, Probe×=1]
→ [RigolDS1000Z_SetTimebase, s/div=0.0001]
→ [RigolDS1000Z_Run]
→ [WaitBlock, 500]
→ [RigolDS1000Z_MathSetup, Operation=FFT, Source1=CH1, FFTWindow=HANNing, FFTUnit=DB, Enable=true]