HP 34401A — 6½-digit Multimeter

Overview

The HP 34401A (also marketed as Agilent 34401A and Keysight 34401A) is a classic 6½-digit bench multimeter capable of DC/AC voltage and current, resistance, frequency, period, diode, and continuity measurements.

Specification Value
Type 6½-digit DMM
Interface GPIB, USB-TMC, RS-232
SCPI command set Yes
Driver name HP34401A
Resource patterns GPIB?*::?*::INSTR, USB?*::?*::INSTR, ASRL?*::INSTR

Connecting

  1. Connect the instrument via GPIB, USB, or RS-232
  2. Click Add Instrument in the sidebar
  3. The Connection Manager scans and lists the resource (e.g. GPIB0::1::INSTR)
  4. Select the resource — the *IDN? response is parsed and the HP34401A driver is highlighted automatically
  5. Click Connect

Front Panel

The HP34401A front panel provides:

Control Description
DCV button Measure DC voltage
ACV button Measure AC voltage (true RMS)
DCI button Measure DC current
ACI button Measure AC current
Ω 2W button Measure resistance (2-wire)
Ω 4W button Measure resistance (4-wire, Kelvin)
Freq button Measure frequency
Period button Measure period
Diode button Diode forward voltage
Cont button Continuity (beeps if below threshold)
Range drop-down Auto or fixed range
Reading display Shows last measured value
Continuous toggle Auto-repeat measurements every second

Sequence Blocks

All HP34401A blocks share a common set of properties:

Property Type Description
Instrument InstrumentSelector The connected HP34401A instance to use
Range Number Measurement range (0 = Auto)
Variable Name Variable Variable that receives the measured value

HP34401A_MeasureDCV

Sends CONF:VOLT:DC {range},DEF then READ?.

Returns voltage in V.

HP34401A_MeasureACV

Sends CONF:VOLT:AC {range},DEF then READ?.

Returns RMS AC voltage in V.

HP34401A_MeasureDCI

Sends CONF:CURR:DC {range},DEF then READ?.

Returns current in A.

HP34401A_MeasureACI

Sends CONF:CURR:AC {range},DEF then READ?.

Returns RMS AC current in A.

HP34401A_MeasureResistance2W

Sends CONF:RES {range},DEF then READ?.

Returns resistance in Ω.

HP34401A_MeasureResistance4W

Sends CONF:FRES {range},DEF then READ?.

Returns resistance in Ω. Use 4-wire for low-resistance measurements to eliminate lead resistance.

HP34401A_MeasureFrequency

Sends CONF:FREQ {range},DEF then READ?.

Returns frequency in Hz.

HP34401A_MeasurePeriod

Sends CONF:PER {range},DEF then READ?.

Returns period in s.


Example Sequence — DC Voltage Log

[Start]
→ [Loop N=60, counter=i]
    ↓ body
    [HP34401A_MeasureDCV, Range=10, Variable=voltage]
    → [AddToChart, Series="Battery Voltage", Variable=voltage]
    → [Wait 1000ms]
→ [SaveCsv, Path="C:\Data\battery_test.csv"]
→ [End]

This sequence measures DC voltage once per second for 60 seconds and saves the results to a CSV file.


SCPI Command Reference

The driver uses the following SCPI commands:

Action Command
Configure DCV CONF:VOLT:DC {range},DEF
Configure ACV CONF:VOLT:AC {range},DEF
Configure DCI CONF:CURR:DC {range},DEF
Configure ACI CONF:CURR:AC {range},DEF
Configure Resistance 2W CONF:RES {range},DEF
Configure Resistance 4W CONF:FRES {range},DEF
Configure Frequency CONF:FREQ {range},DEF
Configure Period CONF:PER {range},DEF
Trigger and read READ?
Identification *IDN?
Reset *RST